# How did FILAB help one of its clients distinguish between the polymer materials polyamide PA 6 and PA 6.6? &#8211; Case study

Case study
# Distinguishing between the polymer materials polyamide PA 6 and PA 6.6

 Our client called on FILAB to address a production discrepancy. One of its machines had been mistakenly fed with the wrong raw material. [Home](https://filab.fr) • [News](https://filab.fr/?post_type=post) • Distinguishing between the polymer materials polyamide PA 6 and PA 6.6
## How did FILAB help a customer distinguish between the polymer materials polyamide PA 6 and PA 6.6?

Polyamide polymers, more commonly known as nylons, are widely used in many industries. PA 6 and PA 6.6 polyamide polymers are the most commonly used. Although they have many similarities, they differ in certain characteristics (mechanical properties, chemical structures, water absorption, and thermal resistance) that make them suitable for specific applications.

This case study focuses on the method and technical means used to determine whether a polyamide is PA6 or PA6.6.

## Summary

 01 Context [](#cont-01) 02 Determination of the chemical nature of parts [](#cont-02) 03 Conclusion of the study [](#cont-03)   01
## Context

Our client called on FILAB to help with a production discrepancy. One of their machines had been mistakenly fed with the wrong raw material. They wanted to take samples from 130,000 parts throughout the production process to determine whether they were made of PA 6.6 (compliant) or PA 6 (non-compliant).

The client commissioned FILAB, a laboratory specializing in material characterization and chemical analysis, to first identify the polymers and then restart production.

   02
## Determination of the chemical nature of parts

What analysis techniques did the laboratory decide to use to distinguish between the two polyamide polymers?

**DSC analysis**
DSC is a thermal analysis technique that measures the differences in heat exchange between a sample to be analyzed and a control (a reference).

In this case, it made it possible to identify the characteristic temperatures of each polymer. The analysis revealed that the sample analyzed was indeed PA 6.6, consistent with normal production.

Pyrolysis GC-MS analysis

This technique confirmed the nature of the polymer found. Pyrolysis coupled with mass spectrometry (Py-GC-MS) is a pyrolysis technique followed by gas chromatography coupled with mass spectrometry (GC-MS) that identifies the organic compounds in a sample and determines its chemical composition by breaking it down into its volatile components.

The nature of the polymer was thus confirmed.

![](https://filab.fr/wp-content/uploads/2024/04/pyroprobe.png)

   04
## Conclusion of the study

FILAB's in-depth analysis made it possible to determine with precision that the parts manufactured and analyzed complied with our client's specifications. This information was essential for successfully restarting production. 

  Need help? Contact our team [](#filab-contact-form)
## Other case studies you may be interested in

### Racket coverings

 Indeed, the intensive use of table tennis rackets has revealed problems related to the durability of this equipment. [Learn more +](https://filab.fr/en/blog/2025/12/degradation-and-optimization-of-the-durability-of-table-tennis-racket-coatings-case-study.md)
### Produits pharmaceutiques

 L’un de nos clients a détecté la présence de particules inconnues dans ses produits pharmaceutiques lors des contrôles de qualité [Learn more +](https://filab.fr/en/blog/2025/12/unknown-particles-in-your-pharmaceutical-products-from-emergency-to-continuous-monitoring.md)
### Rouging phenomenon

 With its excellent properties, stainless steel has become indispensable in many fields: everyday objects, medicine, surgery, automotive, aeronautics, mechanical industries, etc. However, many industrial stainless steel installations are affected by the phenomenon of rouging. [Learn more +](https://filab.fr/en/blog/2025/12/how-did-filab-help-one-of-its-customers-understand-a-rouging-phenomenon-in-stainless-steel.md) [View all our case studies](https://filab.fr/en/blog-filab.md)
## Services related to the case study

 [## Viscosity index measurement on polymer](https://filab.fr/en/our-services/our-analysis-services/polymer-materials-analysis-laboratory/viscosity-index-measurement-on-polymer.md) [## Materials Characterization](https://filab.fr/en/our-services/our-analysis-services/laboratory-material-testing.md) [## Polymer characterization](https://filab.fr/en/our-services/our-analysis-services/polymer-materials-analysis-laboratory.md) [## Polymer characterization laboratory](https://filab.fr/en/our-services/our-analysis-services/laboratory-material-testing/laboratory-polymer-characterization.md) [## Deformulation](https://filab.fr/en/our-services/our-expertise-services/laboratory-deformulation-analysis-reverse-engineering.md)   [![avis filab](https://filab.fr/wp-content/themes/astra-child/templates/assets/img/avis-verifies.svg) 5/5 ![note filab](https://filab.fr/wp-content/themes/astra-child/templates/assets/img/stars.svg)](https://www.avis-verifies.com/avis-clients/filab.fr)
## Laboratory specializing in analysis and expertise for industrial companies

 ![More than 140 people](https://filab.fr/wp-content/uploads/2024/10/collaborateurs-1.svg) More than 140 people at your service ![5200 m² laboratory](https://filab.fr/wp-content/uploads/2024/10/labo.svg) 5200 m² laboratory + 99% of services are provided in-house ![Accredited laboratory](https://filab.fr/wp-content/uploads/2024/10/ISO-17025-1.svg) Accredited laboratory COFRAC ISO 17025 ![+2200 clients](https://filab.fr/wp-content/uploads/2024/10/clients.svg) +2200 clients in France and abroad
### Contact us

 Our services are reserved for professionals.  [![phone](https://filab.fr/wp-content/themes/astra-child/templates/assets/img/form/phone.svg) Call
 us +33 (0)3 80 52 32 05](tel:+33380523205) [![location](https://filab.fr/wp-content/themes/astra-child/templates/assets/img/form/location.svg) Visit us 80 rue Jean-Louis Auguste Petitjean
 21850 - Saint Apollinaire - FRANCE](https://www.google.com/maps/place/Filab/@47.317329,5.076545,15z/data=!4m6!3m5!1s0x47f29e6ab37d6a1d:0xa2cc1016c4f74422!8m2!3d47.317793!4d5.0767354!16s%2Fg%2F1v3gr7f0?entry=ttu) [![mail](https://filab.fr/wp-content/themes/astra-child/templates/assets/img/form/envelope.svg) Contact
 us contact@filab.fr](mailto:contact@filab.fr?subject=%5BVotre%20demande%20:%20How%20did%20FILAB%20help%20one%20of%20its%20customers%20understand%20a%20rouging%20phenomenon%20in%20stainless%20steel?%20en%20%5D) For complaints: all complaints must be sent by mail to contact@filab.fr or by contacting the person who signed the report. The complaint handling procedure is available upon request. In the event of a complaint concerning the laboratory's activities, the customer may be notified of the complaint's handling.