Case study

Distinguishing between the polymer materials polyamide PA 6 and PA 6.6

Our client called on FILAB to address a production discrepancy. One of its machines had been mistakenly fed with the wrong raw material.
Accueil Blog Distinguishing between the polymer materials polyamide PA 6 and PA 6.6

How did FILAB help a customer distinguish between the polymer materials polyamide PA 6 and PA 6.6?

Polyamide polymers, more commonly known as nylons, are widely used in many industries. PA 6 and PA 6.6 polyamide polymers are the most commonly used. Although they have many similarities, they differ in certain characteristics (mechanical properties, chemical structures, water absorption, and thermal resistance) that make them suitable for specific applications.

This case study focuses on the method and technical means used to determine whether a polyamide is PA6 or PA6.6.

Summary

01
Context
02
Determination of the chemical nature of parts
03
Conclusion of the study
01

Context

Our client called on FILAB to help with a production discrepancy. One of their machines had been mistakenly fed with the wrong raw material. They wanted to take samples from 130,000 parts throughout the production process to determine whether they were made of PA 6.6 (compliant) or PA 6 (non-compliant).

The client commissioned FILAB, a laboratory specializing in material characterization and chemical analysis, to first identify the polymers and then restart production.

02

Determination of the chemical nature of parts

What analysis techniques did the laboratory decide to use to distinguish between the two polyamide polymers?

DSC analysis
DSC is a thermal analysis technique that measures the differences in heat exchange between a sample to be analyzed and a control (a reference).

In this case, it made it possible to identify the characteristic temperatures of each polymer. The analyses revealed that the sample analyzed was indeed PA 6.6, consistent with normal production.

Pyrolysis GC-MS analysis

This technique confirmed the nature of the polymer found. Pyrolysis coupled with mass spectrometry (Py-GC-MS) is a pyrolysis technique followed by gas chromatography coupled with mass spectrometry (GC-MS) that identifies the organic compounds in a sample and determines its chemical composition by breaking it down into its volatile components.

The nature of the polymer was thus confirmed.

04

Conclusion of the study

FILAB's in-depth analyses made it possible to determine with precision that the parts manufactured and analyzed complied with our client's specifications. This information was essential for successfully restarting production. 

Need help? Contact our team

Other studies that may be of interest to you

Fracture facies study

Inspect a broken metal anchor to identify the cause of the failure. Find out more

Comparative study of two joints Find out more

Galvanic coupling: securing a naval assembly

One of our customers encountered sealing issues during the final validation tests of their product. Find out more

Services associated with the case study

Viscosity index measurement on polymer

Polymer characterisation

Polymer testing and characterization

Reverse Engineering

Materials Characterization

Laboratory specializing in analysis and expertise for industrial companies

More than 140 people
More than 140 people at your service
5200 m² laboratory
5200 m² laboratory + 99% of services are provided in-house
Accredited laboratory
Accredited laboratory COFRAC ISO 17025
+2200 clients
+2200 clients in France and abroad

Contact us

Our services are reserved for professionals.