Determination of sterilization residues by ETO according to ISO 10993-7

Chemical analysis Problem solving R&D support
More than 140 people
More than 140 people at your service
5200 m² laboratory
5200 m² laboratory + 99% of services are provided in-house

Industrials of the Health sector, you wish to verify the conformity of your devices to ISO 10993-7

What is ISO 10993-7?

Ethylene oxide and ethylene chlorate sterilization are widely used sterilization methods for their microbiological effectiveness. These sterilizations are used in particular for medical devices and pharmaceutical products that cannot withstand traditional high-temperature steam sterilization (such as single-use medical devices for example).

Thus, the ISO 10993-7 standard “Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residues” is an international regulation related to the sterilization of finished products. It allows the determination of residues from ethylene oxide and ethylene chlorate sterilization of medical devices and verifies the maximum elimination of residues of these chemicals.

The implications of ISO 10993-7

Unlike other pollutants, the limits for EtO are non-negotiable. The standard defines maximum permissible levels of two key substances:

  • Ethylene Oxide (EtO): the sterilizing gas itself.
  • Ethylene Chlorhydrine (ECH): a toxic reaction byproduct that forms in the presence of chlorides (often found in PVC).

Our expertise in analysis and desorption

We help you validate your sterilization cycles and desorption (aeration) times through a rigorous analytical protocol.

Determining the exposure category

The standard adjusts the limits based on the duration of contact between the device and the patient:

  • Limited exposure: cumulative contact < 24 hours.
  • Prolonged exposure: contact between 24 hours and 30 days.
  • Continuous exposure: contact > 30 days.

Sensitive analytical techniques

GC-FID: a robust method for quantification

Gas chromatography coupled with flame ionization detection (GC-FID) is particularly well-suited for the analysis of volatile organic compounds such as OTE, CHE, and EG.

It offers:

  • precise and reproducible quantification
  • good sensitivity for volatile compounds
  • analytical robustness suitable for routine analysis
GC-MS: identification and confirmation of compounds

Gas chromatography coupled with mass spectrometry (GC-MS) provides a complementary level of information.

It allows for:

  • formal identification of molecules through their spectral signature
  • confirmation of detected compounds
  • detection of very low trace amounts or unexpected compounds

FILAB supports you in the determination of ethylene oxide according to ISO 10993-7

Why choose FILAB for analysis according to ISO 10993-7

As experts in the analysis of trace organic substances, the FILAB laboratory supports you at every stage of validating your industrial processes, particularly sterilization.

Our specialists have developed specific methods for quantifying Ethylene Oxide and Ethylene Hydrochloride in medical devices. This method involves exhaustive extraction to recover all residues. The analysis is performed using HS/GCMS (Headspace Injection followed by Gas Chromatography coupled with Mass Spectrometry).

Our FAQ

Why should you choose an ISO 10993-7 certified laboratory?

Certification to the ISO 10993-7 standard demonstrates that the laboratory has met the criteria for safety, reliability, and quality in testing medical devices. The certification process ensures that the laboratory has the appropriate equipment, facilities, and personnel to conduct the testing in accordance with the standard's requirements.

In order to be ISO 10993-7 certified, each laboratory must follow the 7 following criteria:

  1. Quality Management System: The laboratory must have a documented quality management system that meets the requirements of ISO 17025 or ISO 9001.
  1. Competence of Personnel: The laboratory must employ personnel who are competent and have the appropriate education, training, and experience to conduct the testing.
  1. Equipment and Facilities: The laboratory must have appropriate equipment and facilities to conduct the testing in accordance with ISO 10993-7.
  1. Test Methods: The laboratory must use test methods that are appropriate for the medical device being tested and in accordance with ISO 10993-7.
  1. Quality Control: The laboratory must have a quality control system in place to ensure the accuracy and reliability of the test results.
  1. Reporting of Results: The laboratory must provide accurate and complete reports of the test results, including the methods used, data obtained, and any limitations or uncertainties associated with the test.
  1. Confidentiality: The laboratory must maintain the confidentiality of all client information, including test results and other proprietary information.
What are the legal requirements regarding ethylene oxide sterilization residuals testing?

The European Union (EU) has specific legal requirements regarding ethylene oxide sterilization residuals testing for medical devices, which are outlined in the Medical Device Regulation (MDR) and In Vitro Diagnostic Regulation (IVDR). These regulations apply to all medical devices sold in the EU, including those that are imported from outside the EU.

According to the MDR and IVDR, manufacturers of medical devices must ensure that their devices are safe and effective. This includes demonstrating that devices have been appropriately sterilized and that any residual ethylene oxide or its by-products do not exceed the maximum allowable limits. The regulations specify the following requirements for ethylene oxide sterilization residuals testing:

  • Biocompatibility Testing: Medical device manufacturers must conduct biocompatibility testing of the device, including testing for ethylene oxide sterilization residuals. The testing must be conducted according to ISO 10993-7.
  • Maximum Allowable Limits: The MDR specifies the maximum allowable limits for ethylene oxide and its by-products, including ethylene chlorohydrin and ethylene glycol. The limits vary depending on the device type and duration of contact with the body.
  • Risk Assessment: Manufacturers must conduct a risk assessment to determine the potential risk to patients associated with any residual ethylene oxide or its by-products.
  • Sterilization Validation: Manufacturers must validate the sterilization process used for their device to ensure that it is effective in eliminating microorganisms and that residual levels of ethylene oxide and its by-products are within the allowable limits.
What are the risks if you do not go through ETO residual testing?

In Europe, medical device manufacturers are required to comply with regulations related to residual testing of ethylene oxide (ETO) as part of the biocompatibility testing of medical devices. If a manufacturer fails to comply with these regulations, there are several legal risks they may face:

  • Regulatory Action: If a medical device manufacturer fails to comply with the ETO residual testing requirements, they may face regulatory action from the competent authorities in Europe. This could include withdrawal of the device from the market, suspension or revocation of the CE marking, and fines.
  • Liability for Product Defects: If a medical device is found to have residual levels of ETO that exceed the allowable limits, the manufacturer may be held liable for any harm caused to patients. The manufacturer could face claims for damages resulting from product defects, including personal injury, loss of income, and other damages.
  • Reputational Damage: Failure to comply with ETO residual testing requirements could damage the reputation of the medical device manufacturer. This could lead to a loss of trust from healthcare professionals and patients and ultimately affect the company's bottom line.
  • Market Access: Medical device manufacturers that fail to comply with European regulations related to residual testing of ETO may be denied market access. This could affect their ability to sell their products in Europe and impact their financial performance.
How to minimize the risks of ethylene oxide sterilization for medical devices?

Ethylene oxide (EO) sterilization is a commonly used method for sterilizing medical devices, but it does come with certain risks. To minimize these risks, it is crucial to implement appropriate measures, including proper validation, monitoring, and residual testing. Here are some key steps to minimize risks associated with ETO sterilization:

  1. Validate the ETO sterilization process and ensure it meets regulatory requirements.
  2. Perform material compatibility studies to ensure devices are compatible with EO.
  3. Properly prepare devices before sterilization, removing sensitive materials.
  4. Allow adequate time for aeration and degassing after ETO exposure.
  5. Conduct routine residual testing using validated analytical methods.
  6. Establish acceptance criteria for ETO residual levels based on safety guidelines.
  7. Document the sterilization process and maintain records for traceability.
Why do I need to perform a toxicological evaluation (ISO 10993-17) if my cytotoxicity tests are compliant?

Cytotoxicity (ISO 10993-5) is a rapid screening test that indicates whether cells die upon contact with the device. However, it does not detect long-term risks such as carcinogenicity, reproductive toxicity, or systemic effects related to the accumulation of small doses of leachable substances. ISO 10993-17 provides rigorous scientific evidence that the released dose is safe for humans over the entire lifespan of the product.

Does ISO 10993-17 apply to all types of medical devices (MD)?

Yes. Although the requirements are more stringent for implants (permanent contact), all medical devices that come into contact with the patient must undergo chemical characterization. If substances migrate to the patient, a toxicological assessment is necessary to determine biosafety, in accordance with the European MDR Regulation 2017/745.

Why is ECH measured in addition to EtO in the ISO 10993-7 standard?

Ethylene Chlorhydrine (ECH) is a reaction byproduct that forms when ethylene oxide comes into contact with chloride ions (found in PVC, some adhesives, or salts). ECH is extremely toxic and often persists longer than EtO in materials. Therefore, monitoring it is crucial for patient safety.

How to get a quote from FILAB?

To obtain a quote, you can contact our team via our contact form, by phone, or by email.

Simply tell us your requirements (type of material, desired analysis, applicable standards, urgency, quantity of samples, etc.). We will then send you a personalized technical and pricing proposal within 24-48 hours.

What is the typical duration of the analysis?

Turnaround times vary depending on the nature of the analysis and the complexity of the expert assessment project.

However, FILAB is committed to providing fast turnaround times tailored to your industrial constraints and urgent needs.

The filab advantages
A highly qualified team
A highly qualified team
Responsiveness in responding to and processing requests
Responsiveness in responding to and processing requests
A COFRAC ISO 17025 accredited laboratory
A COFRAC ISO 17025 accredited laboratory
(Staves available on www.cofrac.com - Accreditation number: 1-1793)
A complete analytical facility of 5,200m²
A complete analytical facility of 5,200m²
Tailor-made support
Tailor-made support
Video debriefing available with the expert
Video debriefing available with the expert
Anaïs DECAUX Customer Support Manager
Ask for your quote